    Name / nom GeCo1
    Absorber / absorbeur

    LUMINEU 14

    Ge wafer - diameter 44 mm - h 175 µm

    Electronic grade - growing axis 100

    SiO coating 70 nm

    Luke grids: GeH + Al anular

    Thermistor

    NTD 34 B serie 2 - flatpack

    5.2 mg

    3 Araldite glue spot and pressure of 25 mg

    Bonding

    NTD: gold wire - 25 µm

    (P = 340 - T = 140)

    Electrodes: Al wire - 25 µm

    (P = 240 - T = 40)

    Supports/ supports teflon supports reduced
    Holder Cu
    Notes Kapton
       

     

    PHOTO

    GeCo.jpg

    GeCo_cialda.png
    CHARACTERIZATION / CARACTERISATION

    1 - Mulet Modane cryostat - semi-vintage electronic - 1/4/2015
    55Fe
    ZnSe: 1 mm x 1 mm x 4 mm
    LED: hamamatsu 8245 - hamamatsu 7850 - without optic fiber

    I bias = 8.32 nA

      V grids = 0 V V grids = 30 V V grids = 70 V V grids = 90 V
    Sensibility [µV/keV] 0.527 4.1 9.3 10.3
    Sigma 55Fe [keV] 0.155+-0.003 1.2+-0.1 2.3+-0.2 3.0+-0.4
    Sigma baseline [eV] 91+-2 11.4+-0.1 8.1+-0.3 7.7+-0.2

     

    Load curves:

    GeCo_VIcurve.png

    R0 [ohm] T0 [K] g0 alpha
    43.1 2.8 7.49 e-6 2.5

     

     

    2 - Ulisse cryostat - semi-vintage electronic - 15/3/2016
    55Fe
    LED: optic fiber 820nm

    the detector is at 22 mK

    M1:    I bias =2.5 nA - Rntd=650 kohm
    M12, M13:    I bias =1.25 nA

    M14:    I bias =5 nA - Rntd=409 kohm

      M1 M12 M13 M14
      V grids = 0 V V grids = 0 V V grids = 0 V V grids = 0 V
    Sensibility [µV/keV] 0.699 0.727 0.738 0.659
    Sigma 55Fe [keV] 0.294 0.741 0.297 0.269
    Sigma baseline [eV] 0.119 0.179 0.153 0.0934